Tool 04
Antenna Near-field Measurement
Plan near-field scan size, sampling interval, probe spacing, and measurement grid requirements for antenna measurement setups.
Parameters
Enter a reference frequency, AUT or scan size, measurement distance, and maximum measurement angle. Set the graph sweep range separately below.
Measurement Geometry
Planar mode uses measurement distance z directly for scan area sizing.
Advanced Settings
The default sample spacing is λref/2.
Sweep Graph
Measurement distance z uses each frequency point's wavelength basis: 5.000λ(f).
Custom Data Step applies to Linear X scale; Log X uses automatic logarithmic points.
Reference Values
- λref
- 29.979 mm
- Measurement z
- 149.896 mm
- Rotation Radius
- 0.000 mm
- Effective z
- 149.896 mm
- Points
- 103 × 103
- Available AUT
- 1.481 × 1.481 m
Diagram
Formulas
Reference Wavelength
Calculate the wavelength from the reference frequency.
Sample Spacing
Planar near-field sample spacing. The default is λref/2.
Measurement Distance
Reference calculations convert distance to a λref multiple; frequency sweeps convert each point to its wavelength multiple.
Cylinder Effective Distance
Cylinder mode adds the AUT rotation radius to the measurement distance before scan area sizing.
Scan Area
Add the margin from effective measurement distance and target angle to the AUT size.
Point Count
Divide scan area by sample spacing to calculate the required point count.
Max AUT Size Inversion
Maximum AUT size that can satisfy the target angle within the given scan area.